Abstract
The small-field radiotherapy is a developing technique because it can reduce the damage to normal tissues. However measurement of the dose distributions of small radiation fields in radiotherapy is a challenge. In this work, we designed and produced an optic fiber X-ray sensor array with high spatial resolution. The sensing array includes 7 sensing probes connecting to an 8-channel optical switch and a photon counting detector (PCD). To verify the practicality of the system, these sensors were measured under a 10×10 cm2 field of a medical linear accelerator. For the small-field application, the dose distribution of radiotherapy fields 1×1 cm2 and 0.8×0.8 cm2 were measured. The distribution of these small radiotherapy fields were given based on the experimental results.
Original language | English |
---|---|
Article number | 127703E |
Journal | Proceedings of SPIE - The International Society for Optical Engineering |
Volume | 12770 |
Issue number | 1 |
DOIs | |
Publication status | Published - 2023 |
Event | Optics in Health Care and Biomedical Optics XIII 2023 - Beijing, China Duration: 14 Oct 2023 → 16 Oct 2023 |
Keywords
- fiber sensors
- optical fiber X-ray sensor array
- optical switch
- small-field radiotherapy