Abstract
Objective: To evaluate the feasibility of a novel “safety-valve” device for preventing catheter related urethral trauma during urethral catheterization (UC). To assess the opinions of clinicians on the performance of the safety-valve device. Materials and Methods: A validated prototype “safety-valve” device for preventing catheter balloon inflation related urethral injuries was prospectively piloted in male patients requiring UC in a tertiary referral teaching hospital (n = 100). The device allows fluid in the catheter system to decant through an activated safety threshold pressure valve if the catheter anchoring balloon is misplaced. Users evaluated the “safety-valve” with an anonymous questionnaire. The primary outcome measurement was prevention of anchoring balloon inflation in the urethra. Secondary outcome measurement was successful inflation of urinary catheter anchoring balloon in the bladder. Results: Patient age was 76 ± 12 years and American Society of Anaesthesiologists grade was 3 ± 1.4. The “safety-valve” was utilized by 34 clinicians and activated in 7% (n = 7/100) patients during attempted UC, indicating that the catheter anchoring balloon was incorrectly positioned in the patient's urethra. In these 7 cases, the catheter was successfully manipulated into the urinary bladder and inflated. 31 of 34 (91%) clinicians completed the questionnaire. Ten percent (n = 3/31) of respondents had previously inflated a urinary catheter anchoring balloon in the urethra and 100% (n = 31) felt that a safety mechanism for preventing balloon inflation in the urethra should be compulsory for all UCs. Conclusion: The safety-valve device piloted in this clinical study offers an effective solution for preventing catheter balloon inflation related urethral injuries.
Original language | English |
---|---|
Pages (from-to) | 179-183 |
Number of pages | 5 |
Journal | Urology |
Volume | 115 |
DOIs | |
Publication status | Published - May 2018 |