This work exploits the high-affinity of recombinant antibodies and low background electrochemiluminescence (ECL) for cardiac-biomarker detection. The developed assay is capable of fg mL-1 detection limits as well as the detection of C-Reactive Protein (CRP) over a clinically relevant range. The assay demonstrated robust reproducibility, selectivity and stability while also highlighting a novel platform for detection of cardiac biomarkers at low concentrations.
Original language | English |
---|---|
Pages (from-to) | 67874-67877 |
Number of pages | 4 |
Journal | RSC Advances |
Volume | 5 |
Issue number | 83 |
DOIs | |