Skip to main navigation Skip to search Skip to main content

In silico modeling of transcatheter heart valve oversizing and ellipticity, Part I: Establishing credibility of an advanced model

  • Sam Boxwell
  • , Dylan Armfield
  • , Rachel M.E. Cahalane
  • , William Hickey
  • , Scott Cook
  • , Patricia Kelly
  • , Philip Cardiff
  • , Laoise M. McNamara
  • University of Galway
  • University College Dublin
  • Harvard University
  • Boston Scientific Corporation

Research output: Contribution to journalArticlepeer-review

Abstract

Background and Objectives Transcatheter aortic valve implantation (TAVI) is the most common modality of treatment for aortic stenosis. However, transcatheter heart valves (THVs) can be prone to early failure and an increase in thrombogenic events, yet the risk factors associated with these failure modes remain poorly understood. Computational modeling may be used to predict biomechanical and hemodynamic indices associated with degeneration and thrombogenicity, however existing models do not fully account for complex stent and leaflet material behavior, and establishing model credibility according to ASME VV-40 is required. Methods In this study, we developed an advanced structural and hemodynamic in silico framework to predict the in vitro performance of a supra-annular, self-expanding THV across a range of clinically-relevant expansion and ellipticity indices. The THV was modelled by incorporating a novel 3-fiber material model for pericardium tissue leaflets and a super-elastic nitinol stent. Results Calculation verification was conducted and, on this basis, we provide recommendations on mesh density, element integration and target time increment. Following verification, we validated our models with radial force, structural high-speed camera and hemodynamic particle image velocimetry testing across multiple THV deployment configurations. In the ‘nominal sizing, circular’ case, we predicted a similar geometric orifice area (4.35 vs 4.02 cm2), pinwheeling index (2.6% vs 2.7%), stent deflection (1.95 vs 1.76 mm) and flow velocity (1.33 vs 1.27 m/s) to in vitro data. Conclusion We validated a novel structural and hemodynamic in silico framework for studying THVs, which will be applied to understand deployment factors contributing to structural degeneration and thrombogenicity. This framework also holds potential for guiding next-generation THV design and predictive procedural modeling.

Original languageEnglish
Article number109404
JournalComputer Methods and Programs in Biomedicine
Volume282
DOIs
Publication statusPublished - 1 Aug 2026
Externally publishedYes

Keywords

  • Computational fluid dynamics
  • Finite element analysis
  • Transcatheter aortic valve implantation
  • Transcatheter heart valve
  • Validation
  • Verification

Fingerprint

Dive into the research topics of 'In silico modeling of transcatheter heart valve oversizing and ellipticity, Part I: Establishing credibility of an advanced model'. Together they form a unique fingerprint.

Cite this