Abstract
Ozone measurement based on the optical absorption of visible light at 603 nm using polymer optical fibres is investigated and tested. Comparisons with a commercial UV based sensor demonstrate the ability of the visible based sensor to measure high concentrations over a range 27 - 127 mg/litre. A resolution of 5 mg/litre is also demonstrated for this sensor.